As follows.
・Measurement of all kinds of mixture of two gases theoreticallypossible.
・Continuous measurement possible.
・Short warm-up time
・No relative gas
・Power saving
・long-lasting sensor
As follows.
・Measurement of all kinds of mixture of two gases theoreticallypossible.
・Continuous measurement possible.
・Short warm-up time
・No relative gas
・Power saving
・long-lasting sensor
As follows.
1.Continuous measurements possible.
2.short warm-up time
3.No relative gas needed
Etc.
Because concentration value is converted from molecular weight, it causes the error when other substances (including water, dust) are contained. Install a filter etc. before INLET of the analyzer to prevent sample gases from mixing with other substances.
Measurement is not possible because the ultrasonic sensor would corrode.
It is possible under following conditions.
1. One gas concentration is fixed.
2. Concentration ratio is constant.
3. When each gas concentration is variable, Measurement can be done by taking in the concentration signal of one of the three gases that is measured and outputted by another gas analyzer.
It depends on the amount and gas composition, approx. delivery times are as follows.
US-100-5VS type.2/US-100-12VS type.2/US-ⅡT-S/US-ⅡT-SH: in a month.
US-ⅡE/US-ⅡT-SR/US-ⅡT-P/US-ⅠT-P: in two months.
Possible. Contact us and let us know your requests.
Because sound is difficult to travel in a high concentration of very light gas (ex. Helium, Hydrogen) such as in vacuum, standard model cannot measure it. So we have specially developed US-ⅡT-SH to measure it.
Sorry, butthere are no models on our line-up.
We can issue it at extra fee.
Check again if the concentration of standard gas is correct and suitable.
After that, calibrate with two gases of zero calibration gas and span calibration gas.
Place the analyzer away from the source of electric noise and pull the ground wire out.
The calibration data of the analyzer can be put back into default status(factory setting).In the case of Ultrasonic module, Calibrate again with two gases of zero calibration gas and span calibration gas.